Technical Information: RADI FAST Hantavirus 6-plex Detection Kit
1. Summary
The RADI FAST Hantavirus 6-plex detection kit is a high-performance real-time PCR (qPCR) solution designed for centralized laboratories and diagnostic centers.
this kit is optimized for standard laboratory workflows, offering exceptional speed and multi-target differentiation.
It provides reliable results in under 50 minutes, significantly increasing laboratory throughput compared to traditional qPCR assays that typically require over 90 minutes.
2. Performance Characteristics
2.1 Analytical Sensitivity (Limit of Detection)
● Limit of Detection (LoD): 10 copies/μL validated across all 6 targets.
● Assay Configuration: The kit utilizes a 2-vial primer/probe system for comprehensive differentiation.
○ Hanta-1 Mixture: Dobrava-Belgrade (DOBV), Seoul (SEOV), and Puumala (PUUV).
○ Hanta-2 Mixture: Sin Nombre (SNV), Andes (ANDV), and Hantaan virus (HTNV).
● High Precision: Optimized for rapid cycling without compromising sensitivity, ensuring accurate detection of low-titer samples.
● Internal Control (IC): Continuous monitoring of the IC validates the integrity of the entire process—from nucleic acid extraction to PCR results.
2.2 Analytical Specificity (Cross-reactivity)
Extensive testing confirms 100% specificity with zero cross-reactivity against clinically relevant pathogens:
| Category | Tested Pathogens (Negative Result) |
| Differential Diagnosis | Leptospira interrogans, Orientia tsutsugamushi, SFTSV |
| Hemorrhagic Fevers | Dengue, Zika, Yellow fever virus |
| Respiratory Viruses | Influenza A/B, Adenovirus |
3. Strategic Advantages
● Ultra-Fast Results: Complete PCR runs in under 50 minutes, compared to the standard 90+ minutes of conventional kits.
● 6-Target Differentiation: Uniquely identifies 6 major Hantaviruses in a single multiplex format.
● Universal Compatibility: Validated for use on major qPCR platforms, including Bio-Rad CFX96, Thermo Fisher Quantastudio 5, and Applied Biosystems (ABI) 7500.
● High Throughput: Optimized for large-scale screening in centralized diagnostic environments.
4. Operating Specifications
| Specimen Type | Serum, Plasma |
| Specimen Volume | 200 μL |
| Storage Condition | -25°C to -15°C |
| Shelf Life | 24 months |
5. Result Interpretation Criteria
● Results are determined based on the Cycle Threshold (Ct) values of the target pathogens and the Internal Control (IC).
| Target Pathogen (Ct ≤ 40) | Internal Control (IC) | Result | Action / Interpretation |
| Detected | +/- (Any) | Positive | Hantavirus detected. IC result is ignored. |
| Not Detected | Detected | Negative | Target not detected. Valid run. |
| Not Detected | Not Detected | Invalid | Test failed. Re-test from extraction. |
| Multiple Detected | +/- (Any) | Invalid | Likely contamination. Re-test required. |
6. Product Composition
Each kit contains reagents for 50 tests:
● Mastermix: High-efficiency 1-vial mastermix.
● Hanta-1 Primer/Probe Mixture: Targets DOBV, SEOV, and PUUV.
● Hanta-2 Primer/Probe Mixture: Targets SNV, ANDV, and HTNV.
● Positive Control: Hantavirus positive control for result validation.
● RNase-free Water: Negative control